CJA
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS

This Article
Right arrow Full Text (PDF)
Right arrow Submit a scholarly reply
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Dube, L.
Right arrow Articles by Blaise, G.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Dube, L.
Right arrow Articles by Blaise, G.

Canadian Journal of Anesthesia, Vol 42, 922-927, Copyright © 1995 by Canadian Anesthesiologists' Society


ARTICLES

Comparison of two administration techniques of inhaled nitric oxide on nitrogen dioxide production

L Dube, M Francoeur, E Troncy, R Carrier and G Blaise
Biomedical Physics Department, Notre-Dame Hospital, University of Montreal, Quebec, Canada.

The purpose of this study was to verify whether, compared with the introduction of the NO-N2 mixture at the air inlet of the ventilator (classical method), a direct injection of NO-N2 into the inspiratory line of the ventilator circuit with a new injection device (new method), would reduce NO2 formation by reducing contact time between O2 and NO. The effect of two FIO2(0.21 and 0.90) and NO concentrations on NO2 production was determined. In the classical method, NO and O2 were mixed with an air/oxygen blender before the gas mixture entered the ventilator. In the new method, NO was injected directly into the respiratory line with the injection system. Nitric oxide and nitrogen dioxide gases were measured using a chemiluminescence analyzer. For a FI02 of 0.90 and 90 ppm of NO2, the amount of NO2 produced was decreased from 8.9 +/- 0.8 ppm (mean +/- SD) with the classical injection system to 4.4 +/- 0.2 ppm with the new injection system (P = 0.0039, Mann-Whitney test), and NO2 production was decreased from 4.5 +/-0.2 ppm to 2.1 +/- 0.4 ppm (P = 0.02) at 60 ppm of NO. However, at a FIO2, no difference was found in the amount of NO2 produced. We conclude that, compared with the classical method of NO administration, the new NO injection system reduces considerably the concentration of inhaled NO2 when a high FIO2 and a high concentration of NO are used.


This article has been cited by other articles:


Home page
Am. J. Respir. Crit. Care Med.Home page
E. TRONCY, J.-P. COLLET, S. SHAPIRO, J.-G. GUIMOND, L. BLAIR, T. DUCRUET, M. CHARBONNEAU, and G. BLAISE
Inhaled Nitric Oxide in Acute Respiratory Distress Syndrome . A Pilot Randomized Controlled Study
Am. J. Respir. Crit. Care Med., May 1, 1997; 157(5): 1483 - 1488.