CJA
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS

This Article
Right arrow Full Text (PDF)
Right arrow Submit a scholarly reply
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Patel, N.
Right arrow Articles by Hancock, D. E.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Patel, N.
Right arrow Articles by Hancock, D. E.

Canadian Journal of Anesthesia, Vol 41, 738-741, Copyright © 1994 by Canadian Anesthesiologists' Society


ARTICLES

The influence of tape type and of skin preparation on the force required to dislodge angiocatheters

N Patel, CE Smith, AC Pinchak and DE Hancock
Department of Anesthesiology, Case Western Reserve University, Cleveland, Ohio.

The study evaluated the effects of different techniques used to secure intravenous (i.v.) catheters. An angiocatheter attached to standard i.v. tubing was taped to human forearm using a standard taping method. A calibrated piezoelectric force transducer was attached to the i.v. tubing. The force applied along the longitudinal axis to pull out the taped catheter was measured and recorded on paper. Three tape types, Curity, Leukopor and Transpore, were evaluated alone, with benzoin skin pretreatment and with mastisol pretreatment. A randomized 3 x 3 block design with 20 replications per block was utilized, and a total of 180 pullout tests were performed on two adult volunteers. Without pretreatment, the forces required to dislodge catheters were (means +/- SEM) 46 +/- 2, 37 +/- 2 and 38 +/- 2 Newtons for Curity, Leukopor and Transpore tape, respectively. Corresponding values for mastisol pretreatment (64 +/- 1, 64 +/- 3 and 52 +/- 3 Newtons) were greater (P < 0.05) for each tape compared with benzoin (54 +/- 3, 53 +/- 2 and 40 +/- 2 Newtons) and no pretreatment. The most frequent failure mode for Transpore tape was by tape fracture, for Curity tape was by separation from the skin of tape and catheter as a single unit, and for Leukopor tape was by catheter separation while tape remained attached to skin (P < 0.001). The data suggest that the application of mastisol prior to taping i.v. catheters with Curity or Leukopor tape helps to minimize the risk of accidental dislodgement.


This article has been cited by other articles:


Home page
Anesth. Analg.Home page
B. A. Williams, M. A. Bolland, S. L. Orebaugh, M. T. Bottegal, and M. L. Kentor
Skin Reactions at the Femoral Perineural Catheter Insertion Site: Retrospective Summary of a Randomized Clinical Trial
Anesth. Analg., May 1, 2007; 104(5): 1309 - 1310.
[Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH TABLE OF CONTENTS
Copyright © 1994 by the Canadian Anesthesiologists' Society.